Anesthetic Monitoring:
Monitoring anesthetic depth
Stage 1: Inducement, excitement, pupils constricted, voluntary struggling
Stage 2: Obtunded reflexes, pupil diameters start to dilate, still excited, involuntary struggling
Stage 3: There are three planes- light, medium, and deep
Stage 4 (mainly dead): Very deep anesthesia. Respiration ceases, cardiovascular function depresses and death ensues immediately.
Due to arrival of newer inhalation anesthetics and concurrent use of injectable anesthetics and neuromuscular blockers the above classic signs do not fit well in most circumstances.
Modern concept has two stages simply dividing it into ‘awake’ and ‘unconscious’.
One should recognize and familiarize the reflexes with different physiologic signs to avoid any untoward side effects and complications
The system must be continuously monitored, and not neglected in favor of other signs of anesthesia.
Take all the information into account, not just one sign of anesthetic depth.
A major problem faced by all anesthetists is to avoid both ‘too light’ anesthesia with the risk of sudden violent movement and the dangerous ‘too deep’ anesthesia stage.
Relating changes of reflexes to anesthetic depth:
Palpebral reflexes: varies between species. Dogs may have no reflex at adequate surgical depth although complete loss in horses indicates moderately deep anesthesia
Corneal reflex: does not disappear until deep anesthesia. Should always be present
Nystagmus: usually indication of excitement and light anesthesia. However, dissociative anesthetics (e.g., Ketamine) cause nystagmus at moderate anesthetic depth. In horses, central stimulation induced by severe hypoxia or hypercapnia also causes this phenomenon, and should not be confused with light plane of anesthesia as animals are perishing.
Lacrimation: Parasympathetic stimulation, usually signs of light plane of anesthesia
Medioventral eye ball position: the most desirable position in most species with the exception in horses (central)
Jaw tone: moderate to loose, most desirable
Gross purposeful movement
Reflex movement in response to stimulation
Immediate hemodynamic response to stimulation; sudden marked increase in heart rate or blood pressure.
Immediate response to stimulation; sudden marked increase in respiratory rate or depth of breathing.
Response to stimulation prior to actual incision (such as clipping, surgical preparation, drape, clamps).
History of vaporizer setting
Muscle tone (e.g. jaw tone)
Pupillary light reflex
Palpebral reflex
corneal reflex
Moist cornea (lacrimination or tears)
Position of the eyeball
Heart Rate: may increase or decrease with increased depth of anesthetic plane. At deep plane it may accompany bradycardia, but also severe cardiovascular depression may increase heart rate to compensate for fall in stroke volume. (differentiate from noxious stimulation induced increase; less marked in changes).
Respiration rate: may increase or decrease with deepening anesthesia. At deep plane, it may accompany apnea, but also may initiate a rapid shallow breath following a period of apnea as a compensation. (similar to hemodynamics, fall in tidal volume tends to increase respiration rate.)
Blood pressure is similar to heart rate, but probably more reliable. In general, the volatile anesthetics; halothane, isoflurane, sevoflurane, and desflurane produce a dose-dependent decrease in arterial blood pressure and many anesthetists use this depression to assess the depth of anesthesia.
The experienced anesthetist relies most of the time on animal’s response to stimuli produced by the surgeon or procedure to indicate adequate depth of unconsciousness.
The most effective depth is taken to be that which obliterates the animal’s response to noxious stimuli without depressing circulatory and respiratory systems.
Anesthetic monitoring of physiologic function:
The primary goal of monitoring anesthetized animals is to ensure adequate tissue perfusion with oxygenated blood.
Monitoring circulation, oxygenation, ventilation and body temperature in the anesthetized patient allows the veterinary anesthetist to identify problems early, institute treatment promptly, and thus avoid irreversible adverse outcomes.
The ACVA published a guideline for standards of monitoring during anesthesia and recovery (JAVMA 1995). The guideline recommends continuous monitoring of the patient by the anesthetist and adequate record keeping of the procedure.
In addition, it recommends patients should be monitored by clinical observations (color, respiratory movement, auscultation etc.), continuous monitoring devices (ECG, BT, pulse oximeter, capnography) and intermittent monitoring devices (Esophageal stethoscope, BP).
Adequate monitoring is needed even for brief anesthetic periods, during the transport of patients, and with sedation that might cause cardiovascular or respiratory complications.
Pulse oximetry, capnography and non-invasive blood pressure monitoring represent three current non-invasive monitoring techniques available to use in the anesthetized patient.
When pulse oximetry and capnography are used together, a beat-to-beat and breath-to-breath non-invasive cardiorespiratory monitoring is provided.
In general, the monitoring physiologic function is divided into subjective or objective monitoring.
Subjective monitoring involves using the anesthetist’s visual, touch, and auditory senses to assess the patient’s vital signs.
Objective monitoring involves monitoring animal’s physiologic function using mechanical devices to amplify the signals that are hard to discern or analyze by human senses, which are then quantified in numeric forms.
Circulation (cardiovascular) monitoring:
Electrocardiography (ECG)
Measures electrical activity of cardiac cells
Other circulatory information including blood pressure, stroke volume and cardiac output is not provided by the ECG.
The ECG leads are positioned usually near the elbow and stifle using a three lead configuration.
Alligator clips are used most commonly to attach the electrode tips to the patient. These can be very traumatic when they are attached for prolonged period of time.
Body position and precise lead placement are not important for monitoring purposes when the primary objective is to describe the electrical pattern in a general manner and monitor for changes that may signal the deterioration of the patient.
In the horse the right arm lead is placed over the heart (right or left side), while the exact location is not important, the left leg and left arm leads are placed in the jugular furrow and the point of the shoulder.
There must be a good contact between the ECG leads and skin, and gels are applied to increase the conduction of the electrical signals.
Alcohol provides for good contact but it evaporates so rapidly as to require frequent application, and is not suitable for long-term monitoring.
Arterial blood pressure:
Cardiac output (CO):
Central venous pressure (CVP):
The goal of monitoring oxygenation in the anesthetized patient is to ensure adequate oxygen concentration in the patient’s arterial blood. To ensure adequate ventilation, the ventilatory function of the anesthetized patient needs to be monitored.
Subjective methods
Clinically, the presence of pink mucous membranes in an anesthetized patient is subjectively indicative of acceptable oxygenation.
However, oxygenation is either difficult or not possible to assess in anemic patients or patients with peripheral vasoconstriction. These patients usually have pale mucous membranes.
Objective methods:
Objective methods include using blood gas analysis for PaO2 (partial pressure of oxygen in the arterial blood) and pulse oximetry.
Pulse oximetry:
Provides a non-invasive, continuous detection of pulsatile arterial blood in the tissue bed, calculates the percentage of oxyhemoglobin present in the arterial blood, and provides the pulse rate of the monitored patient.
Rapidly has become standard care in anesthetized human patients and has gained widespread popularity in veterinary anesthesia.
Methemoglobin and carboxyhemoglobin do not contribute to functional oxygen transport, and one should be aware of the impact of these abnormal hemoglobin species in pulse oximetry.
Abnormal accumulation of methemoglobin species tends to push the oximeter reading toward 85 %, underestimating measurements when SaO2 is above 85 % and overestimating it when below 85%.
Carboxyhemoglobin has light absorption characteristics similar to oxyhemoglobin, and this would contribute to the increase the apparent oxyheoglobin readings in the presence of carboxyhemoglobin.
Affected by motion artifact (e.g, shivering), ambient light, poor peripheral blood flow from hypotension and vasoconstriction, electrical noise from electrocautery, and increased carboxyhemoglobin and methemoglobin levels.
The percentage of hemoglobin saturation with oxygen at different partial pressure of oxygen in the blood is described by the oxyhemoglobin dissociation curve as shown below.
Normal pulse oximeter readings in anesthetized animals should be 99-100%.
Hemoglobin oxygen saturation (SpO2) of 90% corresponds to PaO2 of 60 mmHg which provides definition of hypoxemia if lower than this value. In the clinical setting, PaO2 (as measured by the blood gas analysis) can be estimated using pulse oximetry.
The sites for probe placement include the tongue, ear lip folds, toe pads, axillary or inguinal skin fold, or prepuce/vulva.
The partial pressure of the oxygen can be measured with a blood sample.
Remains the ‘gold standard’ for assessment of oxygenation because of its ability to directly determine the tension of the respiratory gases.
Requires invasive access of blood, does not provide continuous monitoring and is costly
Normal PaO2 in the anesthetized animal that is breathing 100% oxygen should be greater than 200 mmHg (it can be as high as 600 mmHg).
Subjective methods:
May be done by observing chest wall movement or rebreathing bag excursion when the patient is connected to an anesthesia machine.
Auscultation of breathing sound via an esophageal stethoscope or an audible respiratory monitor provides only respiratory rate, and the absence or presence of respiration.
Objective methods:
Requires respirometry, blood gas analysis, or capnography/capnometry.
A respirometer (or respirometry) is an instrument that measures the amount of volume of expired gases.
The device is usually placed between the expiratory limb of an anesthetic machine and the anesthetic breathing hose.
Alternatively, a respirometer connected to a face mask may be used to assess ventilation efficiency in a non-intubated anesthetized patient, although the accuracy is reduced due to air leaks around the mask.
Respirometry assesses tidal volume and minute volume in the anesthetized patient.
Minute volume (VE) is the product of respiratory rate (RR) per minute and tidal volume (VT) of the patient (VE = VT x RR/min).
Respiratory rate and/or tidal volume reduction result in minute volume reduction and reflect the patient’s depressed ventilatory function.
PaCO2, in the patient’s blood can be determined to assess and monitor the anesthetized patient’s ventilation, using a sample collected from the femoral artery or other peripheral artery. Normal values in the anesthetized patient are between 35 mmHg to 45 mmHg.
PaCO2 measurement requires an expensive blood gas analyzer and arterial blood samples may be difficult to collect particularly in small patients. Measuring endtidal CO2 using capnography is a useful alternative to estimate PaCO2 without the need for invasive collection of blood samples
D-E: Inspiratory downstroke, as the patient begins to inhale fresh gas. Phase IV
E-A: Inspiratory pause, where CO2 remains at 0
Urine Output monitoring:
Temperature monitoring:
Patient body temperature should also be monitored during general anesthesia to avoid accidental hypothermia or detect malignant hyperthermia.
Small patients lose body heat very rapidly when anesthetized and precautions should be taken to avoid this.
Body temperature should always be monitored during prolonged surgery of the body cavities. Electronic thermistor probe is commonly placed either in rectum or esophagus for continuous temperature monitoring.
Low body temperature decreases body metabolic rate and impairs the pharmakokinetic profile of drugs on board during anesthesia potentially prolonging recovery.
The temperature should be checked at the end of anesthesia to see whether external heating is required.
The animal in low body temperature will increase muscle contraction to raise body temperature at the time of recovery, a process increasing oxygen demand at the worst time when the tissues need to preserve it most.
Anesthetic record keeping:
In cases where the anesthetic management of a case needs to be defended, an anesthetic record is of enormous worth both as a reminder of the details of the individual cases and as evidence of the general standard of care given by the veterinary practice. To be admissible anesthetic record must be contemporaneous i.e. it must have been made at the same time that the anesthetic was given.
Monitored variables such as heart rate and breathing rate are recorded at regular intervals with minimum of 10 minutes apart, although 5 minute interval recording is a common practice during anesthesia in most veterinary teaching hospitals including Boren Veterinary Medical Teaching Hospital at OSU.
Further readings:
∙ The American College of Veterinary Anesthesiologists guidelines of anesthetic monitoring JAVMA 206 (7) 936-937, 1995
∙ Hall L, Clarke K, and Trim C. Veterinary Anesthesia Saunders 2002
∙ Thurmon J, Benson J, and Tranquilli W Veterinary Anesthesia Williams and Wilkinson 1996
∙ Seymour C and Gleed R (eds) BSAVA Manual of Small Animal Anesthesia and Analgesia 1999